JB6100-A adopts the most advanced Mie scattering principle and convergent Fourier transform optical path in the world. The accuracy and repeatability of the test results are guaranteed by high density probe and full range seamless connection test method. The unique wet circulating dispersion system ensures that there is no deposition of particles in the process of particle testing and no accumulation of waste liquid after drainage, which ensures the accuracy of the second test and makes the test results more reliable. At the same time, the pipeline non-residue design is adopted to ensure the accuracy of testing different samples. The superior optical automatic calibration system shows the unique advantages of JB6100-A.
Brand: Jiuban
Specification type | JB6100-A | |
Executive standard | GB/T 19077-2016/ISO 13320:2009 | |
Test range | 0.01μm -1250μm/0.01μm -2000μm | |
Number of detector channels | 86/96 | |
Accuracy error | <1%(National Standard Sample D50 Value) | |
Repeatability error | <0.5%(National Standard Sample D50 Value) | |
Bubble free | With no exhaust bubbles design, no bubbles interference data more accurate | |
Misoperation protection | The instrument has self-protection function of misoperation and does not respond to misoperation. | |
Camera lens | Imported Canon Lens | |
Laser parameters | Double beam, imported fiber output high power laser lambda = 650nm, P > 10mW | |
Optical path calibration | Automatic Calibration of Optical Path | |
Intelligent operation mode | Software one-key automatic operation | |
Built-in dispersion method | ultrasonic | Frequency: f = 40KHz, power: P = 80W, time: adjustable at will; with ultrasonic anti-dry burning |
Stir | Integrated design of circulating stirring, speed: 100-33950 RPM speed adjustable | |
loop | Circulation rated flow: 0-10L/min adjustable rated power: 25W | |
Sample pool | Self-designed boiling sample cell, better dispersion effect, capacity: 190-600 mL can be tested normally. | |
Micro injection | The instrument can be equipped with micro-automatic testing device, 10 milliliters can be circulated testing (optional) | |
Shading detection | The instrument has the unique function of real-time detection of shading to ensure more accurate sample addition. | |
Dust and Seismic Design | Dust-proof and shock-proof design for the whole machine | |
Anti-corrosion Design (Selection) | It can be equipped with acid resistance, alkali resistance, oil resistance (including all solvent oils), organic solvent resistance (such as acetone, phenol, hexane, isopropanol and all other organic solvents). | |
Testing speed | <1 min/time (without sample dispersion time) | |
volume | 980mm*410mm*450mm | |
weight | 35Kg |